TRRW-9010 – Infant Radiant Warmer
Specifications
- Based on TRRW-9010, TRRW-9018 is added with low-pressure suction S-1, breath resuscitation bag HF-I, oxygen cylinders and infant head fixing unit TJ-I;
- Pre-warm mode, manual mode and baby mode controlled by micro-computer;
- Set temperature and baby temperature can be displayed separately;
- Horizontal angle of warmer module can be adjusted;
- Inclination of the bassinet can be adjusted, and the organic glass panels can be turned outward or disassembled;
- X-ray cassette is under the bassinet;
- APGAR timer;
- Power supply: AC220V-230V/60Hz
- Maximum Power Output: 600W
- Mode of control: Pre-warm mode control, manual mode control and baby mode control by micro-computer
- Baby mode temperature control range: 34.5-37.5°C
- Temperature sensor display range: 5°C-65°C
- Accuracy of skin temperature sensor: 50.3°C
- Temperature uniformity of mattress: 52°C
- Angle of heater head: 0°,30°,60° for two ways;
- Inclination of the bassinet: the three tile position;
- APGAR timer: it will give sound when the unit is tones at 1′,5′, 10′;
- Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on.
Environmental temperature Operating range: 18°C–30°C
- Ambient air movement rate: 50.3m/s